Product Detail

THERMASGARD®RPTM2

Calibratable THERMASGARD® RPTM 2 room pendulum temperature measuring transducer (with globe) THERMASGARD® RPTM 2 with eight switchable measuring ranges, continuous output in an impact resistant plastic housing with quick-locking screws, optionally with/without display, cable sensor with a black plastic globe.
The pendulum sensor has been specially designed to detect the temperature in larger rooms or halls. The resistance thermometre (globe thermometre) achieves a very good, representative measurement result due to its positioning in the room. The dark radiation sensor determines the effective radiation heat at the measured location. This is relevant for calculating the thermal comfort (operative room temperature) taking into account the co-action of thermal radiation and thermal convection. The globe/air temperature ratio is approx. 70%/30%. The sensor is factory-calibrated and can be fine-tuned to ambient conditions by an expert.


No price
Output
Display
  1. Yes
  2. No
Collect
产品编号 1101-1171/1101-1172
  • TECHNICAL DETAILS
  • DOWNLOADS
Power consumption
< 1,0 W/ 24 V DC
< 2,2 VA/ 24 V AC
Temperature deviation
typically ± 0.2 K at +25 °C
Sensor
Pt1000, DIN EN 60751, class B
Connection type
2- or 3-wire connection
Measuring ranges
multi-range switching with 8 switchable measuring ranges
(other ranges optional) Tmin –5 °C, Tmax +60 °C
with manual zero point correction (± 10 K)
Ambient temperature
measuring transducer –5...+60 °C
Globe
plastic, colour black, Ø = 50 mm
Connection cable
PVC; LiYY, 1.5 m
(other lengths optional: e.g. 3 m, 6 m)
Electrical connection
0.14 - 1.5 mm² via terminal screws
Cable connection
cable gland, plastic (M 16 x 1.5; with strain relief, exchangeable, max. inner diameter 10.4 mm) or
M12 connector according to DIN EN 61076-2-101 (optional on request)
Cable gland
M 16 x 1.5, including strain relief, exchangeable,
max. inner diameter 10.4 mm
sensor cable
PVC, H03VV-F, 2 x 0.5 mm²,
KL = approx. 1.5 m (other lengths optional)
Humidity
< 95 % RH, non-precipitating air
Protection class
III (according to EN 60 730)
Protection type
IP 65 (according to EN 60 529) Housing tested,
TÜV SÜD, Report No.713139052 (Tyr 1)
Standards
CE conformity,
electromagnetic compatibility according to EN 61326,
according to EMC Directive 2014/ 30/ EU