AERASGARD® AO2-U (LCD)The measuring device detects the oxygen content O2 (0-25%) in the ambient air. The transmitter converts the measured value into a standardized, analog output signal 0-10 V. As an option, the measuring instrument has a backlit, rotatable display and potential-free changeover contact. The version with backlit LCD display shows the measured values and in/max values. The measuring instrument performs an automatic calibration at regular intervals, which ensures a long-term stable measurement. This can also be performed manually on the instrument at any time. The oxygen meter is optimally suited for use in workplaces, medical facilities, HVAC applications, greenhouses
Collect
Warranty 12 months Voltage supply 24 V DC (± 5%) Power consumption Ø 50mA (depending on backlight, + peak option relay) Operating range, humidity 0...98% r.H., pollutant-free, non-precipitating air Operating range, temperature 0...+50 °C Warm up time 10 min Response time < 10s Output 0 - 10V, 3-wire connection Measuring range 0...25% O2 Accuracy ± 0,2% vol. + max. ±0,5% EW (@ 20°C, 45% r.H., 1013 mbar) Operating temperature 0...+50 °C Storage temperature -20...+50°C Sensor Electrochemical sensor with manual calibration (via zero button) and automatic calibration (fixed) Sensor replacement recommended after 18 months Manual calibration every 3 months, with automatic calibration switched off Electrical connection 0.14 - 1.5 mm2 via terminal screws Housing Polycarbonate with hinge locks Color signal white (similar to RAL 9003) Cable gland M 16 x 1.5; including strain relief Dimensions 90,9 x 80 x 48mm Long-term stability ±1% EW/ Year ±0,2% EW/ Year with auto calibration ON Protection class III (according to EN 60 730) Protection type IP 65 |