Home >> PRODUCTS >>HUMIDITY >>HUMIDITY TRANSMITTERS, HYGROMETERS >> HYGRASGARD® KFF-I (LCD)
Product Detail

HYGRASGARD® KFF-I (LCD)

Calibratable HYGR ASGARD® KFF / KFTF or KFF - 20 / KFTF - 20 duct humidity/temperature transmitters (±2.0% or ±1.8%, respectively) with plastic sinter filter (metal sinter filter optional),housing made of impact resistant plastic, housing cover with quick-locking screws, with/without optional display
The sensor devices measure relative humidity and/or air temperature and convert the measurand into a standard signal of 0-10 V or 4...20 mA. The support four switchable temperature ranges and are applied in non-aggressive dust-free atmospheres in refrigeration, air conditioning, ventilation and cleanroom engineering. Relative humidity (in % r.H.) is the quotient of water vapour partial pressure divided by the saturation vapour pressure at the respective gas temperature. These measuring transducers provide precise detection of humidity. A digital long-term stable sensor is used as measuring element for humidity.The sensor is factory-calibrated and can be fine-tuned to ambient conditions by an expert.


No price
Measured variables
output
Display
  1. Yes
  2. No
output model
Collect
产品编号 1201-3112-0200-029
  • TECHNICAL DETAILS
  • DOWNLOADS
Voltage supply
15...36 V DC
depending on working resistance, residual ripple stabilised ± 0.3 V
Power consumption
< 1.1 VA/ 24 V DC ; < 2.2 VA/ 24 V AC
Working resistance
Ra (ohm) = (Ub -14 V)/ 0.02 A for I variant
Measuring range, humidity
0...100 % r. H.
(output corresponding to 4...20 mA)
Deviation of humidity
typical ± 2 % r. H. (20…80 % r. H. ); at + 25 °C, otherwise ± 3 %
Output, humidity
4...20 mA
Sensor protection
plastic sinter filter, Ø 16 mm, L = 35 mm, exchangeable
(optional metal sinter filter, Ø 16 mm, L = 32 mm)
Process connection
by mounting flange, plastic
(included in the scope of delivery)
Protective tube
PLEUROFORM, material polyamide (PA6), with torsion protection,
Ø 20 mm, NL = 235 mm (optionally 100 mm), vmax = 30 m/s (air)
(option on request in stainless steel V2A (1.4301), Ø 16mm)
Electrical connection
2-, 3- or 4-wire connection
0.14 -1.5 mm ² via terminal screws on circuit board
Cable connection
cable gland, plastic (M 16 x 1.5; with strain relief, exchangeable, max. inner diameter 10.4 mm) or
M12 connector according to DIN EN 61076-2-101 (optional on request)
Housing
plastic, UV-stabilised,
material polyamide, 30 % glass-globe reinforced,
with quick-locking screws
(slotted/ Phillips head - combination),
colour traffic white (similar to RAL 9016),
housing cover for display is transparent !
Housing dimensions
72 x 64 x 43.3 mm (Tyr 1 with Display)
Cable gland
M 16 x 1.5, including strain relief, exchangeable,
max. inner diameter 10.4 mm
Long-term stability
± 1 % per year
Permissible air humidity
< 95 % RH, non-precipitating air
Protection class
III (according to EN 60 730)
Protection type
IP 65 (according to EN 60 529) Housing tested,
TÜV SÜD, Report No.713139052 (Tyr 1)
Standards
CE conformity,
according to EMC directive 2014/ 30/ EU,
according to EN 61326-1, according to EN 61326-2-3
Display
two-line display with illumination, 36 x15 mm (W x H),
for displaying ACTUAL temperature and/ or ACTUAL humidity


  • Telephone

    • 19836508612
    • 19836508612
Technical Support: 97建站 | Admin Login
×
seo seo