Home >> PRODUCTS >>HUMIDITY >>HUMIDITY TRANSMITTERS, HYGROMETERS >> HYGRASGARD® KFTF-U NTC10K
Product Detail

HYGRASGARD® KFTF-U NTC10K

Calibratable HYGRASGARD® KFTF (± 2,0%) or KFTF - 20 (± 1,8%) duct humidity/temperature transmitter (±2,0% or ± 1,8%, respectively) with plastic sinter filter (metal sinter filter optional), housing made of impact resistant plastic with quick-locking screws, optionally with/without display, with cable gland (optionally with M12 connector according to DIN EN 61076-2-101).
The sensor device measures relative humidity and/or air temperature and converts the measurands into a standard signal of 0-10 V or 4...20 mA. It supports four switchable temperature ranges and is applied in non-aggressive dust-free atmospheres in refrigeration, air conditioning, ventilation and cleanroom engineering. The measuring transducer provides precise detection of humidity. A digital long-term stable sensor is used as measuring element for humidity. The sensor is factory-calibrated and can be fine-tuned to ambient conditions by an expert.


No price
Collect
产品编号 1201-3111-2015-029
  • TECHNICAL DETAILS
  • DOWNLOADS
Voltage supply
24 V AC (± 20 %) and
15…36 V DC
Power consumption
< 1.1 VA/ 24 V DC ; < 2.2 VA/ 24 V AC
Measuring range, humidity
0...100 % r. H.
(output corresponding to 0 -10 V)
Deviation of humidity
typical ± 2 % r. H. (20…80 % r. H. ); at + 25 °C, otherwise ± 3 %
Output, humidity
0 - 10 V
Measuring range, temperature
multi-range switching
with 4 switchable measuring ranges
–35...+35 °C; –35...+75 °C; 0...+50 °C; 0...+80 °C
(output corresponding to 0 -10 V or 4...20 mA)
Temperature deviation
typically ± 0.2 K at +25 °C
Output, temperature
0 - 10 V
Sensor protection
plastic sinter filter, Ø 16 mm, L = 35 mm, exchangeable
(optional metal sinter filter, Ø 16 mm, L = 32 mm)
Ambient temperature
storage –35...+85 °C;
operation – 30...+75 °C, non-precipitating
Process connection
by mounting flange, plastic
(included in the scope of delivery)
Protective tube
PLEUROFORM, material polyamide (PA6), with torsion protection,
Ø 20 mm, NL = 235 mm (optionally 100 mm), vmax = 30 m/s (air)
(option on request in stainless steel V2A (1.4301), Ø 16mm)
Electrical connection
2-, 3- or 4-wire connection
0.14 -1.5 mm ² via terminal screws on circuit board
Cable connection
cable gland, plastic (M 16 x 1.5; with strain relief, exchangeable, max. inner diameter 10.4 mm) or
M12 connector according to DIN EN 61076-2-101 (optional on request)
Housing
plastic, UV-stabilised,
material polyamide, 30 % glass-globe reinforced,
with quick-locking screws
(slotted/ Phillips head - combination),
colour traffic white (similar to RAL 9016)
Housing dimensions
72 x 64 x 37.8 (51.8) mm (Tyr 1)
Cable gland
M 16 x 1.5, including strain relief, exchangeable,
max. inner diameter 10.4 mm
Long-term stability
± 1 % per year
Permissible air humidity
< 95 % RH, non-precipitating air
Protection class
III (according to EN 60 730)
Protection type
IP 65 (according to EN 60 529) Housing tested,
TÜV SÜD, Report No.713139052 (Tyr 1)
Standards
CE conformity,
according to EMC directive 2014/ 30/ EU,
according to EN 61326-1, according to EN 61326-2-3


  • Telephone